Procedure for RapidFit™
Standard Lamination Adapter Procedure For Transtibial
andTransfemoral Laminated Sockets.
(RapidFit # 2022)
Swanson Regional Orthotic & Prosthetic Research Center has developed
the
RapidFit Method to improve patient care for lower
limb amputees.
RapidFit
Overview
Before you begin: Read
directions completely
RapidFit #2022 has slots for Carbon Tape
Reinforcement
Materials Needed: RapidFit
Standard Lamination Adapter # 2022, Mold Sealer,
Dacron Felt, Nyglass and Perlon Stockinettes, PVA Bag
and Acrylic Resin.
Tools Needed: 5mm Hex Key, Cast
Saw, Channel Lock Pliers
Plaster Model Preparation
- Take the cast over the
suction liner and or cast sock.
- Fill the cast with plaster
keeping the pipe 2" from the bottom of the
cast.
- Seal the plaster model using
an appropriate mold sealer such as cellulose
crystals dissolved in acetone.
- Tie a knot in the distal end
of a non-stick nylon stockinette. Apply the
non-stick nylon stockinette to the plaster model
for a vacuum air wick. The knot should be centered
on the distal end of the plaster model.
- Paint the distal end of the
cast around the knot in the nylon with a mold
sealer such as cellulose crystals dissolved in
acetone. This will glue down the nylon for
trimming. Let dry and trim the nylon in a circular
shape slightly smaller than the lock mechanism.
- Apply the PVA bag and cap the
distal end.
- Cut a round piece of stretch
tape the same size as the attachment tooling and
apply to the PVA over the drilled hole. The tape
will help prevent the PVA from tearing. Puncture a
hole in the tape over the drilled center hole.
Fasten the attachment tooling with the four long
nails provided with the RapidFit Kit. NOTE: Using
the longer nails provided by the RapidFit Kit will
provide a more secure attachment.
- Lay-Up Considerations for
Carbon Tape Reinforcement Careful consideration
must be given to choice of materials used in
construction of the laminated socket. The
Prosthetist should consider the patient's age,
weight, height, activity level and the other
components used in the prosthesis. 8) Important:
Use the number of layers of one inch carbon tape
reinforcement according to the patient's K Level
as shown: Level K-1 low activity thru Level K-2
low activity use one layer of carbon tape Level
K-2 moderate activity thru Level K-3 moderate
activity use two layers of carbon tape Level K-3
high activity thru Level K-4 high activity use
three layers of carbon tape
- Apply the 1" carbon tape
reinforcement through the anchor holes in the
RapidFit Adapter according to the above K levels.
- Lubricate the four size M6 x
25mm socket head cap screws for easy removal.
Attach the four-hole 3/8" square spacer block
to the RapidFit™ Lock Adapter with the four
socket head cap screws. Lubricate the M6 X 25mm
socket head cap screw and install in the side hole
in the square spacer block. Tighten all screws
with 5mm hex key. Fill the screw heads with
fitting gel or clay putty.
- Lay-Up and Lamination Careful
consideration must be given to choice of materials
used in construction of the laminated socket. The
Prosthetist should consider the patient's age,
weight, height, activity level and the other
components used in the prosthesis. The following
is an example of a standard lay-up used for an
average patient.
Lay-Up Example
- Apply one layer of ½ ounce
Dacron Felt.. punch a hole in the Dacron the same
size as the attachment screw(s). Be careful not to
cut the PVA bag.
- Cut a round section of ½
ounce Dacron Felt about 9" in diameter. Trim
out a circular hole in the Dacron slightly smaller
in size as the nailed round attachment tooling.
Apply this round Dacron Felt on top of the first
layer of felt. Place the hole over the attachment
tooling. This is needed to hide the carbon tape on
the inside of the socket.
- Apply the Rapidfit Adapter
with 1" carbon tape reinforced anchor holes
according to the K levels to the plaster model.
- Fasten the RapidFit Lock
Adapter to the plaster model using the socket head
cap attachment screw. The Prosthetist should check
and adjust the RapidFit position for correct
rotation, then tighten with hex key. Fill the
screw head with fitting gel or clay putty.
- Apply two layers of Nyglass
stockinette. Tie the Nyglass into the tie ring and
reflex back over the model. Note: Spectralon™
stockinette may be substituted for Nyglass.
- Apply two additional layers
of Nyglass stockinette. Tie the Nyglass into the
tie ring and reflex back over the model. Note:
Spectralon™ stockinette may be substituted for
Nyglass.
- Apply two additional layers
of Nyglass stockinette. Tie the Nyglass into the
tie ring and reflex back over the model. Note:
Spectralon™ stockinette may be substituted for
Nyglass.
- Apply one layer of Perlon
stockinette. Note: Other finish stockinette may be
substituted for Perlon.
- Apply the outer PVA bag and
pour cone. Laminate with acrylic resin such as
Epox-Acryl. Tie off the excess resin with a
string.
Shell Lamination for a
cosmetic cover
- Fill any undercuts with clay
putty or fitting gel. Apply one cotton or nylon
stockinette for a separator. Apply the inner PVA
bag and tie off the top. Apply two nylon
stockinette twisted and reflected back. Apply the
PVA Bag and pour cone. Laminate with 80-20
polyester resin. Trimming 22) Let the lamination
cool down. Sand away the material covering the end
of the lamination over the attachment screw.
Remove the socket head cap screw with a hex key.
- Sand away the material
covering the end of the lamination over the five
socket head cap screws. Remove the four size M6 x
25 mm socket head cap screws with a 5mm hex key.
- Trim out the four-hole square
spacer block. Remove the four-hole square spacer
block. Use the side M6 x 25mm socket head cap
screw to pull out the square spacer block or use
channel lock pliers.
- Trim out and remove the
socket from the plaster model. Sand and buff trim
lines.
- Clean off any excess plastic
on the RapidFit Adapter.
Assembly
- Install the other
endoskeletal components according to
manufacturer's directions. NOTE: A set of four 6
mm x 16 mm stainless steel flat head socket screws
are provided for your convenience. Apply thread
lock such as Loctite 242 to the four 6mm x 16 mm
stainless steel flat head socket screws and
tighten to a torque setting of 12 Nm or 9 foot
pounds.
|