RapidFit™  12 Hole Lock Adapter Lamination Kit For Transtibial and Transfemoral Sockets  (RapidFit # 5040)
Swanson Regional Orthotic & Prosthetic Research Center has developed the
RapidFit Method to improve patient care for lower limb amputees.

RapidFit Overview                                      For a printable version in Microsoft Word select here

Before you begin: Read directions completely

RapidFit # 5040 has slots for Carbon Tape reinforcement

Materials Needed: Lock, RapidFit Adapter # 5040, Mold Sealer, Dacron Felt, Nyglass and Perlon Stockinettes, PVA Bag and Acrylic Resin.

Tools Needed: 5mm Hex Key, #2 Phillips Screwdriver, Standard Screwdriver, Cast Saw, Channel Lock Pliers

Plaster Model Preparation 

  1. Take the cast over the suspension liner.
  2. Fill the cast with plaster keeping the pipe 2" from the bottom of the cast. The plaster model will have a center protrusion replicating the distal end of the suspension liner.
  3. Drill a ¼" hole straight down through the center protrusion 1 1/4" deep.
  4. Flatten the distal end of the model by removing plaster until the diameter of the distal end of the model is equal to the outer dimension of the attachment tooling. Caution: DO NOT exceed the outer dimension of the tooling, because the socket would be too short. If too much plaster was removed, add back plaster to correct.
  5. Check how the tooling is centered on distal end of the model. Any voids between the model and the tooling should be filled with plaster to create a smooth transition.
  6. Seal the plaster model using an appropriate mold sealer.
  7. Apply a non-stick nylon stockinette to the plaster model for a vacuum air wick.
  8. Paint the distal end of the cast around the knot in the nylon with a mold sealer. This will glue down the nylon for trimming. Let dry and trim the nylon in a circular shape slightly smaller than the lock mechanism.
  9. Apply the PVA bag and cap the distal end.
  10. Cut a round piece of stretch tape the same size as the attachment tooling and apply to the PVA over the drilled hole. The tape will help prevent the PVA from tearing. Puncture a hole in the tape over the drilled center hole. Fasten the attachment tooling with the four long nails provided with the RapidFit Kit. NOTE: Using the longer nails provided by the RapidFit Kit will provide a more secure attachment.

    Lay-Up Considerations for Carbon Tape Reinforcement Careful consideration must be given to choice of materials used in construction of the laminated socket. The Prosthetist should consider the patient's age, weight, height, activity level and the other components used in the prosthesis.
  11. Apply the 1" carbon tape reinforcement through anchor slots in the RapidFit Adapter. Apply additional layers of carbon tape as needed according to the patient's activity level.

Lock Preparation 

  1. Install the Center Hole Sleeve to the bottom of the Adapter with Pin Tool. Read the lock manufacturer's directions for preparing the lock for attachment to the plaster model. Remove the lock cartridge from the lock mechanism and fill the interior of the lock body with fitting gel or stick wax to prevent plastic from leaking into the lock. Install and tighten the lock fabrication sleeve with a screwdriver.
  2. Attach the lock to the RapidFit Lock Adapter with the six screws. Move the Carbon tape to the sides to make room for the lock mechanism in the middle of the adapter. Note: The 6 x 3/8" stainless steel sheet metal screws provided are for use with Delrin lock bodies. Use a #2 Phillips Screwdriver to tighten the sheet metal screws.
  3. Attach the four-hole fabrication spacer block to the bottom of the RapidFit Lock Adapter with the four 6 mm socket head cap screws. Tighten the four screws to the spacer block with a 6mm hex key.
  4. Lubricate the four size M6 x 25mm socket head cap screws for easy removal. Attach the four-hole square spacer block to the RapidFit™ Lock Adapter with the four socket head cap screws. Tighten all screws with 5mm hex key Fill the screw heads, set screw for lock cartridge and slot in the fabrication sleeve with fitting gel or clay putty. Also fill in the two side holes in the square spacer block with fitting gel or putty (these threaded holes will help pull the spacer block out)

Lay-Up and Lamination 

Careful consideration must be given to choice of materials used in construction of the laminated socket. The Prosthetist should consider the patient's age, weight, height, activity level and the other components used in the prosthesis. The following is an example of a lay-up used for an average patient.

Lay-Up Example: 

  1. Apply one layer of ½ ounce Dacron Felt. Trim out a circular hole in the Dacron the same size as the nailed round holding spacer. Be careful not to cut the PVA bag.
  2. Cut a round section of ½ ounce Dacron Felt about 12" in diameter. Trim out a circular hole in the Dacron slightly smaller in size as the nailed round attachment tooling. Apply this round Dacron Felt on top of the first layer of felt. Place the hole over the attachment tooling. This is needed to hide the carbon tape on the inside of the socket.
  3. Apply the RapidFit Adapter and lock with 1" carbon tape reinforced anchor holes to the plaster model.
  4. Fasten the Lock Mechanism with the RapidFit Lock Adapter to the plaster model using the socket head cap screw provided with the lock mechanism. The Prosthetist should check and adjust the RapidFit position for correct rotation, then tighten with hex key. Fill the screw head with fitting gel or clay putty.
  5. Apply two layers of Nyglass stockinette. Tie the Nyglass into the tie ring and reflex back over the model. Note: Spectralon™ stockinette may be substituted for Nyglass. 
  6. Apply two additional layers of Nyglass stockinette. Tie the Nyglass into the tie ring and reflex back over the model. Note: Spectralon™ stockinette may be substituted for Nyglass.
  7. Apply two additional layers of Nyglass stockinette. Tie the Nyglass into the tie ring and reflex back over the model. Note: Spectralon™ stockinette may be substituted for Nyglass.
  8. Apply one layer of Perlon stockinette. Note: other finish stockinette may be substituted for Perlon.
  9. Apply the outer PVA bag and pour cone. Laminate with acrylic resin such as Epox-Acryl.

Shell Lamination for a cosmetic cover 

  1. Fill any undercuts with clay putty or fitting gel. Apply one cotton or nylon stockinette for a separator. Apply the inner PVA bag and tie off the top. Apply two nylon stockinette twisted and reflected back. Apply the PVA Bag and pour cone. Laminate with 80-20 polyester resin.

Trimming 

  1. Let the lamination cool down. Sand away the material covering the end of the lamination over the lock screw. Remove the lock socket head cap screw with a hex key.
  2. Sand away the material covering the end of the lamination over the five socket head cap screws. Remove the four size M6 x 25 mm socket head cap screws with a 5mm hex key.
  3. Trim out the four-hole square spacer block. Remove the four-hole square spacer block. Use the side M6 x 25mm socket head cap screw to pull out the square spacer block or use channel lock pliers.
  4. Trim out and remove the socket from the plaster model. Sand and buff trim lines.
  5. Sand off the plastic covering the round fabrication sleeve and remove it with a standard screwdriver.
  6. Clean off any excess plastic on the RapidFit Adapter.

Assembly

  1. Remove any fitting gel or stick wax residue from the lock body and threaded attachment holes. Follow the lock manufacturer's directions to reassemble the lock.
  2. Optional: To install a RapidFit Pin Cover remove the Center Hole Sleeve from the bottom of the Lock Adapter center hole using the Pin Tool. Install a RapidFit Pin Cover or suction accessory insert to the bottom of the RapidFit Adapter center threaded hole. Tighten with pin tool.
  3. Install the other endoskeletal components according to manufacturer's directions. NOTE: A set of four 6 mm x 16 mm stainless steel flat head socket screws are provided with the RapidFit Adapter. Caution, care must be taken to use the correct length flat head screws provided with the RapidFit Adapter so that the screws tighten properly and securely. Do not substitute these flat head screws. Apply thread lock such as Loctite 242 to the four 6mm x 16 mm stainless steel flat head socket screws and tighten to a torque setting of 12 Nm or 9 foot pounds.
 